Use, Maintenance, and Care of the Devices
Gaging Connectors
3-
8
Gaging Procedure
NOTE
Always hold a connector gage by the gage barrel, below the dial indicator. This gives
the best stability, and improves measurement accuracy.
1.
Select the proper gage for your connector. Refer to
for gage part number.
2.
Inspect and clean the gage, gage master, and device to be gaged. Refer to
“Cleaning Connectors” on page 3-4
of this chapter.
3.
Zero the connector gage (refer to
a.
While holding the gage by the barrel, and without turning the gage or the device, connect the
gage master to the gage. Refer to
for more information. Connect
the nut finger tight. Do not overtighten.
b.
Using an open-end wrench to keep the gage from rotating, use the torque wrench
recommended for use with the kit to tighten the gage master connector nut to the specified
torque. Refer to
“Final Connection Using a Torque Wrench” on page 3-13
for more
information.
c.
The gage pointer should line up exactly with the zero mark on the gage. If not, loosen the dial
lock screw on the gage and rotate the gage dial so that the pointer is aligned with the zero
mark.
d.
Tighten the dial lock screw and remove the gage master.
e.
Attach and torque the gage master to the gage once again to verify that the setting is
repeatable. Remove the gage master.
4.
Gage the device connector (refer to
):
a.
If gaging one of the 50
Ω
loads in the kit, use a collet extractor tool to remove the center
conductor collet from the device connector to be gaged. Refer to
for the
part number of this extractor tool. Refer to
mm connector and the location of the center conductor collet.
b.
While holding the gage by the barrel, and without turning the gage or the device, connect the
gage master to the device. Refer to
the nut finger-tight. Do not overtighten.
c.
Using an open-end wrench to keep the gage from rotating, use the torque wrench
recommended for use with the kit to tighten the device connector nut to the specified torque.
Refer to
“Final Connection Using a Torque Wrench” on page 3-13
d.
Gently tap the barrel of the gage with your finger to settle the gage reading.
e.
Read the gage indicator dial. If the needle has moved clockwise, the center conductor is
protruding
by and amount indicated by the
black
numbers. If the needle has moved
counterclockwise, the center conductor is
recessed
by an amount indicated by the
red
numbers.
For maximum accuracy, measure the connector a minimum of three times and take an
average of the readings. After each measurement, rotate the gage a quarter-turn to reduce
measurement variations that result from the gage or the connector face not being exactly
Содержание 85050B
Страница 4: ...4 85050B ...
Страница 7: ...85050B 1 1 1 General Information ...
Страница 13: ...85050B 2 1 2 Specifications ...
Страница 16: ...2 4 85050B Specifications Mechanical Characteristics Figure 2 1 Connector Pin Depth ...
Страница 20: ...2 8 85050B Specifications Electrical Specifications ...
Страница 21: ...85050B 3 1 3 Use Maintenance and Care of the Devices ...
Страница 30: ...85050B Use Maintenance and Care of the Devices Gaging Connectors 3 10 Figure 3 3 Gaging 7 mm Connectors ...
Страница 38: ...85050B Use Maintenance and Care of the Devices Handling and Storage 3 18 ...
Страница 39: ...85050B 4 1 4 Performance Verification ...
Страница 42: ...4 4 85050B Performance Verification Recertification ...
Страница 43: ...85050B 5 1 5 Troubleshooting ...
Страница 47: ...85050B 6 1 6 Replaceable Parts ...
Страница 50: ...6 4 85050B Replaceable Parts Introduction ...
Страница 51: ...85050B A 1 A Standard Definitions ...