2-2
85050B
Specifications
Environmental Requirements
Environmental Requirements
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating
temperature is a critical factor in performance. During a measurement calibration, the
temperature of the calibration devices must be stable and within the range shown in
.
IMPORTANT
Avoid unnecessary handling of the devices during calibration because your
fingers are a heat source.
Table 2-1 Environmental Requirements
Parameter
Limits
Temperature
Operating
a
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
+20
°
C to +26
°
C
Storage
−
40
°
C to +75
°
C
Error-corrected range
b
b. The allowable network analyzer ambient temperature drift during measurement calibration
and during measurements when the network analyzer error correction is turned on. Also, the
range over which the network analyzer maintains its specified performance while correction
is turned on.
±
1
°
C of measurement calibration temperature
Relative humidity
Type tested, 0% to 95% at 40
°
C, non-condensing
Содержание 85050B
Страница 4: ...4 85050B ...
Страница 7: ...85050B 1 1 1 General Information ...
Страница 13: ...85050B 2 1 2 Specifications ...
Страница 16: ...2 4 85050B Specifications Mechanical Characteristics Figure 2 1 Connector Pin Depth ...
Страница 20: ...2 8 85050B Specifications Electrical Specifications ...
Страница 21: ...85050B 3 1 3 Use Maintenance and Care of the Devices ...
Страница 30: ...85050B Use Maintenance and Care of the Devices Gaging Connectors 3 10 Figure 3 3 Gaging 7 mm Connectors ...
Страница 38: ...85050B Use Maintenance and Care of the Devices Handling and Storage 3 18 ...
Страница 39: ...85050B 4 1 4 Performance Verification ...
Страница 42: ...4 4 85050B Performance Verification Recertification ...
Страница 43: ...85050B 5 1 5 Troubleshooting ...
Страница 47: ...85050B 6 1 6 Replaceable Parts ...
Страница 50: ...6 4 85050B Replaceable Parts Introduction ...
Страница 51: ...85050B A 1 A Standard Definitions ...