2-18
General Information
Cleaning Connections for Accurate Measurements
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections
on the cable end face can be detected as well as cracks or chips in the fiber
itself. Use a microscope (100X to 200X magnification) to inspect the entire
end face for contamination, raised metal, or dents in the metal as well as any
other imperfections. Inspect the fiber for cracks and chips. Visible imperfec-
tions not touching the fiber core may not affect performance (unless the
imperfections keep the fibers from contacting).
W A R N I N G
Always remove both ends of fiber-optic cables from any instrument,
system, or device before visually inspecting the fiber ends. Disable all
optical sources before disconnecting fiber-optic cables. Failure to do
so may result in permanent injury to your eyes.
Cleaning Connectors
The procedures in this section provide the proper steps for cleaning fiber-
optic cables and Agilent Technologies universal adapters. The initial cleaning,
using the alcohol as a solvent, gently removes any grit and oil. If a caked-on
layer of material is still present, (this can happen if the beryllium-copper sides
of the ferrule retainer get scraped and deposited on the end of the fiber during
insertion of the cable), a second cleaning should be performed. It is not
uncommon for a cable or connector to require more than one cleaning.
C A U T I O N
Agilent Technologies strongly recommends that index matching compounds
not
be applied to their instruments and accessories. Some compounds, such as
gels, may be difficult to remove and can contain damaging particulates. If you
think the use of such compounds is necessary, refer to the compound
manufacturer for information on application and cleaning procedures.
Table 2-3. Cleaning Accessories
Item
Agilent Part Number
Pure isopropyl alcohol
—
Cotton swabs
8520-0023
Small foam swabs
9300-1223
Compressed dust remover (non-residue)
8500-5262
Содержание 83400 Series
Страница 2: ...Agilent 83400 Series Lightwave Source and Receiver Modules User s Guide ...
Страница 6: ...v Laser Classification IEC and FDA labels and their placement on an Agilent 83402C ...
Страница 9: ......
Страница 11: ......
Страница 17: ...1 6 Getting Started Connecting a Source Figure 1 2 Location of laser safety labels ...
Страница 47: ...2 24 General Information Theory of Operation Figure 2 13 Agilent 83410C and 83412B block diagram ...
Страница 66: ...3 15 Specifications and Regulatory Information Regulatory Information Declaration of conformity sources ...
Страница 67: ...3 16 Specifications and Regulatory Information Regulatory Information Declaration of conformity receivers ...
Страница 81: ...Index 4 Index width spectral procedure 4 4 Y YAG system dual hetrodyne 4 7 ...