Cleaning Information
Preserving Connectors
96
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition
Preserving Connectors
Listed below are some hints on how best to keep your connectors in
the best possible condition.
Making Connections
Before you make any connection you must ensure that all cables and
connectors are clean. If they are dirty, use the appropriate cleaning
procedure.
When inserting the ferrule of a patchcord into a connector or an
adapter, make sure that the fiber end does not touch the outside of the
mating connector or adapter. Otherwise you will rub the fiber end
against an unsuitable surface, producing scratches and dirt deposits
on the surface of your fiber.
Dust Caps and Shutter Caps
Be careful when replacing dust caps after use. Do not press the bottom
of the cap onto the fiber as any dust in the cap can scratch or dirty
your fiber surface.
When you have finished cleaning, put the dust cap back on, or close
the shutter cap if the equipment is not going to be used immediately.
Always keep the caps on the equipment when it is not in use.
All of Agilent Technologies’ lightwave instruments and accessories are
shipped with either laser shutter caps or dust caps. If you need
additional or replacement dust caps, contact your nearest Agilent
Technologies Sales/Service Office.
Immersion Oil and Other Index
Matching Compounds
Wherever possible, do not use immersion oil or other index matching
compounds with your device. They are liable to impair and dirty the
surface of the device. In addition, the characteristics of your device
can be changed and your measurement results affected.
Cleaning Instrument Housings
Use a dry and very soft cotton tissue to clean the instrument housing
and the keypad. Do not open the instruments as there is a danger of
electric shock, or electrostatic discharge. Opening the instrument can
cause damage to sensitive components, and in addition your warranty
will be voided.
Содержание 81618A
Страница 9: ...Table of Contents 8 Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition ...
Страница 16: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 15 Accessories ...
Страница 22: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 21 Specifications ...
Страница 36: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 35 Performance Tests ...
Страница 88: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 87 Cleaning Information ...
Страница 111: ...110 Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition ...
Страница 112: ......
Страница 113: ... Agilent Technologies GmbH 2001 2002 Printed in Germany February 20 2002 81623 90B12 81623 90B12 www agilent com ...