Performance Tests
Performance Tests
50
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition
Change Setup
53
Disable Atty1 and switch the output with the monitor output.
54
Set the attenuation of Atty1 to 35 dB and of Atty2 to 25 dB.
55
Enable Atty1 again.
56
Adjust the attenuation of both attenuators in the following order:
• Atty1: DUT Power Meter shows a reading of -37.2 dBm.
• Atty2: the REF Power Meter shows a reading of -47.2 dBm.
-40 dBm Range (equivalent to -10 dBm range for 81628B)
57
Disable Atty1.
58
Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
59
Enable Atty1.
60
Switch one range up to the -30 dBm (0 dBm for the 81628B) range
and note the power readings (#15).
61
Switch one range down (-40 dBm, -10 dBm for the 81628B) and note
the power readings again (#16).
62
Increase the attenuation of Atty1 by 10 dB and note the results in
the test record (#17).
63
On the DUT switch one range down to the -50 dBm range (-20 dBm
for the 81628B).
64
If necessary, adjust the attenuation of Atty1 in order to be on the
upper limit of the range (i.e. -x7.y dBm).
65
Decrease the attenuation of Atty2 by 10 dB in order to be on the
upper limit of the -50dBm range at the reference power meter.
-50 dBm Range (equivalent to -20 dBm range for 81628B)
66
Disable Atty1.
67
Zero both power meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
68
Enable Atty1.
69
Switch one range up to the -40 dBm (-10 dBm for the 81628B) range
and note the power readings (#18).
70
Switch one range down (-50 dBm, -20 dBm for the 81628B)) and
note the power readings (#19).
Содержание 81618A
Страница 9: ...Table of Contents 8 Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition ...
Страница 16: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 15 Accessories ...
Страница 22: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 21 Specifications ...
Страница 36: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 35 Performance Tests ...
Страница 88: ...Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition 87 Cleaning Information ...
Страница 111: ...110 Agilent 81618A 9A and Agilent 81622B 3B 4B 6B 7B 8B User s Guide Second Edition ...
Страница 112: ......
Страница 113: ... Agilent Technologies GmbH 2001 2002 Printed in Germany February 20 2002 81623 90B12 81623 90B12 www agilent com ...