5 Putting ParBERT into Operation
How to Use the Agilent 81250 Config Tool
5-14
Agilent 81250 ParBERT Installation Guide, May 2004
How to Specify Startup Settings
The
Startup Settings
define how the ParBERT user software is started.
Figure 32 Startup Settings Page
When you are setting up a new ParBERT and its controller PC, it is
recommended that you keep the defaults.
Operating Modes
Local mode is used, if the system is operated from the ParBERT
controller PC. Local mode starts both the Agilent 81250 firmware server
and the user interface.
Controlled mode is used, if the ParBERT controller PC shall be operated
through GPIB or LAN. It starts the Agilent 81250 firmware server and the
GPIB Gateway, not the user interface. If you do not use the GPIB
Gateway, you can disable it.
NOTE
Local mode
Controlled mode
Содержание 81250
Страница 1: ...Agilent Parallel Bit Error Ratio Tester 81250 ParBERT Installation Guide Agilent Technologies ...
Страница 12: ...1 About this Manual Document History 1 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Страница 30: ...2 Introduction Unpacking ParBERT 2 18 Agilent 81250 ParBERT Installation Guide May 2004 ...
Страница 138: ...Appendix Declaration of Conformity 10 2 Agilent 81250 ParBERT Installation Guide May 2004 Declaration of Conformity ...
Страница 139: ...Declaration of Conformity Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 3 ...
Страница 141: ...Site Attenuation Requirements Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 5 ...
Страница 142: ...Appendix Site Attenuation Requirements 10 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Страница 145: ......
Страница 146: ...Copyright Agilent Technologies 2004 Printed in Germany May 2004 81250 91010 Agilent Technologies ...