Setting up a proprietary System
Setting up an Extended Agilent 81200 System based on Agilent E4849B/C and Agilent E4848B Mainframes
29
Clock Reference Interconnection
Tools Required
You need a flat screw driver and a TORX No. 362 screw driver. A TORX screw driver is attached to
the shipment.
Steps to perform
Each of the mainframes is equipped with an Agilent E4805A Central Clock Module. Use the delivered
clock distribution cables (p/n E4805-61601, the cables are already connected to the clock module in
the Extender mainframes) to connect from one Agilent E4805A module in one mainframe to the other
Agilent E4805A module in the other mainframe. Up to two additional mainframes can be connected.
It is necessary to unplug the Agilent E4805A Central Clock Module from the Master mainframe, open
the side panel and connect the flat cable from Extender 1 and Extender 2 into the connectors on the
main board of the module.
1. In the Master mainframe, disconnect all clock distribution cables to the Agilent E4841A or
E4861A modules going from the Central Clock module of the Master mainframe.
Figure 20
Disconnect all Clock Distribution Cables from the Clock Module
2. Unscrew the two screws, on top and bottom of the front panel of the Agilent E4805A Clock
Module, which secure the module to the mainframe. A flat screw driver is required.
3. Remove the Agilent E4805A Clock Module from the Master mainframe.
Disconnect all Clock Cables
from the Agilent E4805A Central Clock Module
This ends of the clock cables can be kept connected
Содержание 81200
Страница 1: ...Agilent 81200 System Installation Guide ...
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 8: ...8 ...
Страница 12: ...Getting Started Steps to Get Started 12 ...
Страница 36: ...Setting up a proprietary System Display Control Panel 36 ...
Страница 50: ...Start the System How to Test your System 50 ...
Страница 75: ......
Страница 76: ... Copyright Agilent Technologies 1999 2000 Edition E0800 Printed in Germany E4849 91023 ...