Installation
8-11
Install Standard NEMA Switch Wall Boxes
It is the customer’s responsibility to supply and install standard NEMA
switch wall boxes with conduit knockouts within the local distribution
cable length limitations of where the instrument site will be.
Note
Minimum depth of the wall box must be at least 7 cm (2 3/4 in.).
When installing XSDC branch cable runs, use dual-gang wall boxes to
facilitate connector installation.
Agilent Technologies will supply and install faceplates onto NEMA wall
boxes. Faceplates are included in each SDN connector kit.
Standard single-gang SDN/UTP faceplate, 1 hole, 78599AI-J12
Standard dual-gang SDN/UTP faceplate, 1 hole, 78599Al-J14
Note
The use of non-standard wall boxes must be approved by Agilent
Technologies prior to installation. When non-standard wall boxes are used,
the customer must supply the associated faceplates with the appropriate size
connector hole.
Install Branch Cables
Agilent Technologies supplies three different spool lengths of SDC branch
cable and two different spool lengths of XSDC branch cable. The customer
must determine the length of each branch cable run and add
approximately 10% overrun. Order the associated spools of branch cable to
accommodate your branch cable requirements. Refer to the following
spool listing and associated part numbers.
The customer is responsible for installing each branch cable run from the
ACC to each wall box. Label both ends of each branch cable with an
appropriate destination l.D. (bed number, room number).
Branch cable runs must be continuous; do not splice cables or mix cable
types.
Caution
Only branch cable supplied by Agilent Technologies may be used to
interconnect the component parts of the SDN. No other type or brand
of cable may be substituted, or the warranty may be void.
There are three types of branch cables used to interconnect the ACC to the
wall box; unshielded twisted pair (UTP), the standard system distribution
cable (SDC), and extended system distribution cable (XSDC). Refer to
“SDN Installation Restrictions and Limitations” on page 8-4.