Shutdowns
Shutdown 4—Front inlet flow shutdown
4 of 10
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
❐
The column is plugged or not installed correctly.
Check the column and the column installation.
❐
The inlet’s pressure sensor is not operating correctly.
If you are using a split/splitless inlet:
❐
The split ratio is too low. Increase the amount of split flow.
❐
The inlet’s proportional control valve is stuck open or closed because of
contamination or another fault.
If you are using a purged packed or cool on-column inlet:
❐
The inlet’s control valve is stuck closed because of contamination or
another fault.
Shutdown 4—Front inlet flow shutdown
This message indicates the front inlet failed to reach its flow setpoint in the
allotted time or cannot maintain its flow setpoint. When you are operating in
flow-control mode, the inlet has 2 minutes to reach the setpoint before
shutdown. The instrument will be not ready until the problem is corrected
and the inlet reaches the flow setpoint.
See Shutdown 3—Front inlet pressure shutdown for a list of potential causes
and solutions.
Shutdown 5—Back inlet pressure shutdown
See Shutdown 3—Front inlet pressure shutdown for a list of potential causes
and solutions.
Shutdown 6—Back inlet flow shutdown
See Shutdown 4—Front inlet flow shutdown for a list of potential causes and
solutions.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......