Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Shutdowns
840
Shutdown 3—Front inlet pressure shutdown
Shutdown 3—Front inlet pressure shutdown
This message indicates the inlet failed to reach its setpoint in the allotted
time or cannot maintain its pressure setpoint. The inlet’s flows shut off (split/
splitless inlets go to purge mode) and the instrument goes to a not ready state
until the problem is corrected and the inlet reaches the pressure setpoint.
For EPC inlets, the allotted time to reach the pressure setpoint before the
inlet shuts down varies with the type of inlet as shown in the following table.
The following may be causing the error:
❐
The initial gas supply pressure is too low to reach the setpoint.
Make sure that the initial gas supplies are high enough to support the
pressure and flow setpoints. The pressure at the initial supply should be
at least 5 psi (10 psi for the split/splitless inlet) greater than the desired
setpoint.
❐
A large leak is present somewhere in the system.
Use an electronic leak detector to find leaks; correct them. Refer to
chapter 200 for the leak test procedure for your inlet. In addition to
checking the gas supply connections and the inlet, remember that the
column could be broken, which would cause a large leak.
❐
If you are using the gas saver, make sure that the gas saver flow rate is
high enough to maintain the highest column-head pressure during a run.
❐
The flow is too low for the column in use.
Make sure that the flow is adequate for the column you are using.
Type of inlet
Lapse in time before shutdown
Purged packed, cool on-column
2 minutes
Split/splitless
5.5 minutes
Auxiliary
4 minutes
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......