Installing a Split/Splitless Inlet
Restore the GC to operating condition

11.

If you 

do not

 have a gas or liquid sample valve

—Connect the insert 

weldment to the inlet body and tighten the large nut with the inlet 
wrench.

If you 

do

 have a gas or liquid sample valve

—Connect the insert 

weldment attached to the valve to the inlet body and tighten the large 
nut with the inlet wrench. Connect the gang block fitting attached to the 
valve to the front of the flow module.

Restore the GC to operating condition

1.

Install the capillary adapter, if used.

2.

Restore the column connection.

3.

Restore carrier and other gases to the instrument.

4.

Restore power.

5.

Apply your normal operating pressures. Leak-check the manifold, back 
panel, and column fittings.

Содержание 6850 Series II

Страница 1: ...isassemble it during installation Description Quantity T 20 Torx screws M4 x 8 mm 2 Machine screws M4 x 0 7 12 mm 6 Bottom cup insulation 1 Middle cup insulation 1 Top cup insulation 1 Insulation cup...

Страница 2: ...ttom block insulation Top block insulation Screws Split Splitless capillary inlet system 1 2 3 4 Disposable wrist strap Bottom cup insulation Middle cup insulation Top cup insulation 1 Flow Module 2 S...

Страница 3: ...nch T 20 Torx screwdriver Overview Caution Before starting review the safety information listed at the end of this document 1 Disconnect the GC 2 Remove the lid top cover 3 Remove the existing inlet i...

Страница 4: ...otentially dangerous shock hazard by disconnecting the power cord before working on the instrument 1 Turn off the GC and unplug the power cord Allow time for all heated zones to cool 2 Open the lid If...

Страница 5: ...t If the GC currently has an inlet installed you must remove it 1 Loosen the screws holding the connector cover plate next to the inlet flow module Lid top cover without valve box accessory Remove T 2...

Страница 6: ...ssary you can also slide the insulation sleeve off of the bottom of the inlet Purged Packed inlet 1 Trace the heater sensor cable from the inlet to the wiring harness connector Disconnect it 2 Remove...

Страница 7: ...art of the valve assembly Move it out of the way Disconnect the gang block fitting also part of the valve assembly from the flow module Remove the flow module 3 Use forceps to remove the liner and O r...

Страница 8: ...o the sides of the lid 3 Place the flow module in position and secure it with three screws 4 Align the inlet as shown Lower it into the block insulation 5 Secure the inlet with three screws 6 Remove t...

Страница 9: ...w module cable into its connector Replace the connector cover plate and tighten the screws 9 Connect the carrier supply line to the back of the flow module 10 Connect the heater sensor cable to the wi...

Страница 10: ...ect the insert weldment attached to the valve to the inlet body and tighten the large nut with the inlet wrench Connect the gang block fitting attached to the valve to the front of the flow module Res...

Страница 11: ...Installing a Split Splitless Inlet Restore the GC to operating condition 11...

Страница 12: ...attention to a condition or possible situation that could A caution calls attention to a condition or possible situation that could On the instrument all phases of operation service and repair of this...

Отзывы: