Installing a Split/Splitless Inlet
Restore the GC to operating condition
11.
If you
do not
have a gas or liquid sample valve
—Connect the insert
weldment to the inlet body and tighten the large nut with the inlet
wrench.
If you
do
have a gas or liquid sample valve
—Connect the insert
weldment attached to the valve to the inlet body and tighten the large
nut with the inlet wrench. Connect the gang block fitting attached to the
valve to the front of the flow module.
Restore the GC to operating condition
1.
Install the capillary adapter, if used.
2.
Restore the column connection.
3.
Restore carrier and other gases to the instrument.
4.
Restore power.
5.
Apply your normal operating pressures. Leak-check the manifold, back
panel, and column fittings.