
Installation Guide: Flame Ionization Detector (FID)
15
3
From inside the oven, place the cup over the detector fitting
so its top edge touches the top of the oven.
4
Release the spring while making certain it fits into the groove
in the detector fitting.
Restoring the GC to Operating Condition
1
Restore gas supplies and check for leaks, particularly at all
new connections. Follow the leak test procedures described
in your 6820
Maintenance and Troubleshooting
manual.
2
Reinstall GC covers and panels.
3
Restore GC electrical power.
4
Switch off electrical power, then switch it on again to ensure
the new configuration is properly retained in GC memory.