6
Operating in CI Mode
CI Autotune
146
5977B Series MSD Operating Manual
Table 11 Reagent gas settings
Reagent gas
Methane
Isobutane
Ammonia
EI
Ion polarity
Positive
Negative
Positive
Negative
Positive
Negative
N/A
*
* N/A Not Available
Emission 150
A
50
A
150
A
50
A
150
A
50
A
35
A
Electron energy
150 eV
150 eV
150 eV
150 eV
150 eV
150 eV
70 eV
Filament
1
1
1
1
1
1
1 or 2
Repeller
3 V
3 V
3 V
3 V
3 V
3 V
30 V
Ion focus
130 V
130 V
130 V
130 V
130 V
130 V
90 V
Entrance lens offset
20 V
20 V
20 V
20 V
20 V
20 V
25 V
EM volts
1200
1400
1200
1400
1200
1400
1300
Shutoff valve
Open
Open
Open
Open
Open
Open
Closed
Gas select
A
A
B
B
B
B
None
Suggested flow
20%
40%
20%
40%
20%
40%
N/A
Source temp
250 °C
150 °C
250 °C
150 °C
250 °C
150 °C
230 °C
Quad temp
150 °C
150 °C
150 °C
150 °C
150 °C
150 °C
150 °C
Interface temp
280 °C
280 °C
280 °C
280 °C
280 °C
280 °C
280 °C
Autotune
Yes
Yes
No
Yes
No
Yes
Yes
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...