54845-68803

ATX Service Kit 
User Instructions

Description

The 54845-68803 ATX Service Kit contains a fan safety shield for 
covering the exposed fan/s during servicing with the sleeves removed. 
A service disk is also included and provides software for accessing the 
oscilloscope operating system or resetting the attenuator contact 
counter.

The 54845-68803 ATX Service Kit contains the following items:

Fan Safety Shield Instructions

Follow these instructions to install and use the Fan Safety Shield.

W A R N I N G

Risk of Injury.

 Use caution when working around the Infiniium cooling fan/

s with the instrument sleeves removed. The cooling fan/s,  exposed on one side, 
can be hazardous. When the cabinet sleeves are removed, install the 
Fan Safety Shield supplied with this kit to protect your fingers from the 
fan blades.

Agilent Part Number

Description

Quantity

54810-00601

Fan Safety Shield

1

54810-68760

Infiniium Service Disk Pouch

1

5061-7354

Fan Bracket Tool

1

54845-92001

These User Instructions

1

s1

Содержание 54845-68803

Страница 1: ...s Fan Safety Shield Instructions Follow these instructions to install and use the Fan Safety Shield W A R N I N G Risk of Injury Use caution when working around the Infiniium cooling fan s with the instrumentsleeves removed The cooling fan s exposed on one side can be hazardous When the cabinet sleeves are removed install the Fan Safety Shield supplied with this kit to protect your fingers from th...

Страница 2: ...om cover If necessary refer to the Replacing Assemblies procedures in the Infiniium 54845B 46B Oscilloscope Service Manual 4 Clipthefanshield over the outsideofthechassisasshowninthepicture below 5 You may now service the ATX Oscilloscope with power applied 6 You must remove safety shield before re installing sleeve and bottom cover The safety shield contains a raised ridge to prevent against this...

Страница 3: ... removal and installation of the ATX frame fan brackets Removing Fan Bracket 1 Insert the assistance tool as shown in the picture below 2 With tool in place pry upwards to remove the bracket Installing Fan Bracket 1 Using thumb push bracket up against chassis 2 Insert assistance tool as shown below ...

Страница 4: ...4 Fan Bracket Assistance Tool 3 While prying down use thumb to push bracket into position ...

Страница 5: ...niium oscilloscope 2 Connect a mouse and a keyboard to the oscilloscope 3 Insert the service disk in the floppy drive and turn on the oscilloscope 4 The oscilloscope will boot into the Windows operating system 5 PressControl Esc on the keyboard to access the Windows Start menu 6 Highlight Programs select Accessories and then select System Tools Click the Scan Disk icon 7 In the ScanDisk applicatio...

Страница 6: ...nnect a mouse and a keyboard to the oscilloscope 3 Insert the service disk in the floppy drive and turn on the oscilloscope 4 The oscilloscope will boot into the Windows operating system 5 PressControl Esc on the keyboard to access the Windows Start menu 6 Highlight Programs and click the Infiniium Scope icon ...

Страница 7: ...the Windows operating system 5 PressControl Esc on the keyboard to access the Windows Start menu 6 Open the Programs group and click on the Infiniium Scope icon 7 When the oscilloscope application comes up click on the mouse icon in the upper right corner of the screen 8 Click on the Utility pull down menu click on Selftest In the Selftest screen click on Select Service Extensions 9 Click on the S...

Страница 8: ...nel you are changing 3 Under Set Value enter the corrected attenuator value using the keyboard or the drop down keypad then press the button labeled Apply to Set Value to Actuation Count 4 Repeat for each channel you need to change Agilent Technologies Printed in Malaysia Agilent Part Number 54845 92001 Printed January 2002 ...

Отзывы: