Operation
48
Agilent 5100 and 5110 ICP-OES User’s Guide
2
Select the features and accessories on the Configuration page to
be used during analysis. (Some features are only available in ICP
Expert Pro software.)
3
On the ‘Elements’ page, select the element(s) from the ‘Element’
drop-down box or type the element name or symbol and then
perform one of the following:
Click to add the primary wavelength for the selected
element.
Highlight the wavelength you wish to use from the list of
available wavelengths displayed. Click
Add.
NOTE
Alternatively, press CTRL and select the element from the periodic table to add
the primary wavelength for the selected element to your method.
The element will appear in the table with the selected wavelength
and default settings selected.
4
Check that there are no known interferents or other analytical
lines close to the selected analytical line. Their relative intensity
will govern how close the lines can be to each other. For example,
if your matrix contains an element that is not of analytical
interest, but is a potential interferent and has a line close to one
of your analytical lines, the concentration of that element in your
matrix will determine whether you need to choose another
analytical line.
5
Make any required adjustments to each element including
selecting a different wavelength, entering additional information
into the ‘Label’ column, and selecting the type (choose from
analyte, internal standard or interferent).
6
Click
Conditions
to modify both common settings for the run and
settings for each element. Up to eight different measurement
conditions can be used.
7
Click
QC
to enter the method detection limits, select the QC tests
to be used and which error actions should be performed if an
error occurs.
8
Click
IEC
to enter the concentration of the elements in your
analyte and interferent standards.