Safety Practices and Hazards
12
Agilent 5100 and 5110 ICP-OES User’s Guide
The acid concentration in the sample that is measured varies,
depending upon the digestion steps and acid types used. Instrument
users should be aware of the hazards associated with use of the acids
used for sample preparation and apply all necessary precautions
including use of lab coats, safety goggles and other appropriate forms
of personal protection. The acid wastes should be disposed of in
accordance with local regulatory requirements.
The type, volatility and concentration of the organic solvents used in
the sample that is measured varies, depending upon the selected
solvent and the sample preparation involved. Instrument users
should be aware of the hazards associated with use of the organic
solvents for sample preparation, and apply all necessary precautions
including ensuring adequate ventilation during use, and use of lab
coats, safety goggles, gloves and other appropriate forms of personal
protection. The organic wastes should be disposed of in accordance
with local regulatory requirements.
Air flow to the air intake port of the spectrometer and accessories
must be unobstructed. Do not block the ventilation grills on the
spectrometer and accessories. Consult the manuals supplied with
your PC, monitor, printer and water-cooling system for their specific
ventilation requirements.
Great care should be taken when working with glass or quartz parts
to prevent breakage and cuts. This is especially important when
inserting the nebulizer into the spray chamber, or removing and
replacing pieces of broken torch.
The spectrometer weighs approximately 106 kg (234 lb). To avoid
injury to personnel or damage to the instrument or property, always
use suitable mechanical lifting device to move the instrument.
Use only Agilent-supplied or approved spares with your instrument.
Only trained operators should use the instrument.
Содержание 5100 ICP-OES
Страница 1: ...Agilent 5100 and 5110 ICP OES User s Guide ...
Страница 24: ...Introduction 24 Agilent 5100 and 5110 ICP OES User s Guide This page is intentionally left blank ...
Страница 54: ...Operation 54 Agilent 5100 and 5110 ICP OES User s Guide This page is intentionally left blank ...
Страница 65: ......