
Chapter 10
167
Command Reference
:CREJ:LIM
1
0
. C
o
mm
and
Refe
re
nc
e
:CREJ:LIM
Syntax
[:SENSe][:FIMPedance]:CREJect:LIMit <numeric>
[:SENSe][:FIMPedance]:CREJect:LIMit
Description
Sets the boundary value (percentage of the measurement range) within the detection range
of Low C when you turn on the Low C reject function. The measurement range to which
the set value is applied differs depending on the setup of the measurement range mode as
shown below.
•
When in auto-range mode (ON has been specified with the
command):
When the measurement frequency is in the 1 kHz: 100E-12 F (100 pF) range
When the measurement frequency is in the 1 MHz: 1E-12 F (1 pF) range
•
When in the fixed range mode (OFF has been specified with the
command):
The selected measurement range
For example, if you make a measurement with the range fixed to the 1
µ
F range and specify
1%, Low C is detected if the measured value of the primary parameter (Cs or Cp) is 10 nF
or less.
For information on the screen display, GPIB output, and so on when Low C is detected,
refer to Appendix E, “At-a-glance Table of Operations When Overload or Low C is
Detected,”.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Related commands
Equivalent key
sequence
[Low C Rej]
(
[Blue]
-
[Level]
) -
Limit
<numeric>
Description
Boundary value
Range
0 to 10
Initial value
0
Unit
% (percent)
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...