184
Chapter 10
Command Reference
:RANG
:RANG
Syntax
[:SENSe][:FIMPedance]:RANGe[:UPPer] <numeric>[PF|P|NF|N|UF|U|MF|M|F]
[:SENSe][:FIMPedance]:RANGe[:UPPer]?
Description
Sets the measurement range.
With the measurement frequency and the measurement range set to 1 MHz and 47E-12
(47p) or less, respectively, changing the measurement frequency to 1 kHz will
automatically set the measurement range to 100E-12 (100p). On the other hand, with the
measurement frequency and the measurement range set to 1 kHz and 2.2E-9 (2.2n) or
more, respectively, changing the measurement frequency to 1 MHz will automatically set
the measurement range to 1E-9 (1n).
When you set the measurement range with this command, the measurement range mode is
automatically set to the hold range (specified to OFF with the
command).
Parameters
If one of the settable values is not specified for the parameter, a suitable measurement
range whose recommended range includes the specified parameter (for example, 4.7E-9 if
the specified parameter is 5E-9) is set.
MAX or MIN can be used to specify the parameters.
Query response
{1E-12|2.2E-12|4.7E-12|100E-12|220E-12|470E-12|1E-9|2.2E-9|4.7E-9|10E-9|22E-9|47E-
9|100E-9|220E-9|470E-9|1E-6|2.2E-6|4.7E-6|10E-6}<newline><^END>
If the measurement range mode is Auto, the measurement range used in the immediately
preceding measurement is read out as the query response.
Related commands
Equivalent key
sequence
[Range Setup]
(
[Blue]
-
[Auto/Hold]
)
<numeric>
Description
Measurement range
Range
When the measurement frequency is 1 kHz:
100E-12(100PF), 220E-12(220PF), 470E-12(470PF),
1E-9(1NF), 2.2E-9(2.2NF), 4.7E-9(4.7NF),
10E-9(10NF), 22E-9(22NF), 47E-9(47NF), 100E-9(100NF),
220E-9(220NF), 470E-9(470NF),
1E-6(1UF), 2.2E-6(2.2UF), 4.7E-6(4.7UF),
10E-6(10UF) (only the 16 points listed above).
When the measurement frequency is 1 MHz:
1E-12(1PF), 2.2E-12(2.2PF), 4.7E-12(4.7PF), 10E-12(10PF),
22E-12(22PF), 47E-12(47PF), 100E-12(100PF),
220E-12(220PF), 470E-12(470PF), 1E-9(1NF) (only the 10
points listed above).
Initial value
10E-6
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...