Command Reference
:RANG:AUTO
1
0
. C
o
mm
and
Refe
re
nc
e
:RANG:AUTO
Syntax
[:SENSe][:FIMPedance]:RANGe:AUTO {ON|OFF|1|0}
[:SENSe][:FIMPedance]:RANGe:AUTO?
Description
Selects the measurement range mode from the auto range (automatic range switching) or
hold range (fixed range).
Setting the measurement range (set with the
command) automatically selects the
hold range mode.
Parameters
Query response
{1|0}<newline><^END>
Related commands
Equivalent key
sequence
[Auto/Hold]
Description
ON or 1 (initial value)
Specifies the auto range mode.
OFF or 0
Specifies the hold range mode.