Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
3-53
Measurement Functions
Measurement Ranging Mode
Measurement Ranging Mode
Before executing measurements, you select a ranging mode from the following four
modes. You can set the ranging mode for each measurement unit.
•
•
•
•
The following table lists the allowable measurement ranging modes for each
measurement mode.
Table 3-5
Allowable Measurement Ranging Modes
If you choose sweep measurement or sampling measurement (initial interval
³
2 ms)
and you do not set a ranging mode, auto ranging is set for voltage measurement
mode channel and limited auto (1nA) ranging is set for current measurement mode
channel automatically.
For sampling measurement (initial interval < 2 ms) , an error occurs if fixed range is
not set.
Measurement Mode
Ranging Mode
Sweep Measurement
Auto, Limited Auto, Fixed
Sampling Measurement
(Initial Interval
a
³
2 ms)
a. You specify initial interval on the MEASURE: SAMPLING SETUP
screen.
Auto, Limited Auto, Fixed
Sampling Measurement
(Initial Interval
a
< 2 ms)
Fixed
Knob Sweep
Compliance
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...