3488A-19

Caution:  

The Agilent 3488A mainframe, as well as the plug-in option cards, are susceptible to static

discharges. Use the following anti-static procedures to prevent static damage

.

Static Handling

 

Static electricity is a familiar phenomenon which, except for an occasional shock,

doesn’t seem very serious. However, it has been proven that in the electronics industry electrostatics
discharge (ESD) is a major cause of component failure. In many cases, the component damaged may not
immediately fail, causing low instrument reliability and unnecessary repairs at a later time. ESD damage
can occur at static levels too low for human perception. Additionally, it has been shown that ESD can
affect both passive and active devices.

Caution

:  The HP 3488A mainframe, the plug-in cards, and any assembly or subassembly removed

from the mainframe, must be handled in a static safe environment

.

The following guidelines are the minimum requirements for a static safe service environment.

-

 

The work bench should be equipped with a conductive table mat. The mat should be grounded to
earth ground through a 1 megohm resistor.  The mat should be equipped with at least one swivel
connected to the swivel connector for connecting wrist straps.

-

 

All service and handling personnel should wear a conductive wrist strap in contact with bare skin.
This strap should be connected to the swivel connector on the conductive table mat through a 1
megohm resistor.

-

 

All metal equipment at a work station must be grounded. This includes soldering irons, solder
suckers, shelving, and equipment stands.

-

 

Only one common ground should be provided at a workstation.

-

 

The workstation must be kept free of nonconductors. No common plastics, polybags, cardboard,
cigarette or candy wrappers should be allowed. There should not be rugs or carpet on the floor,
shelving, or bench top.

-

 

Only proper containers should be used for shipping, storing and transporting assemblies.

HP 3488A Upgrade Procedures

The following steps show you how to upgrade an old 3488A mainframe:

STEP 1: Cover Removal

It is necessary to remove the top cover to access any of the major assemblies within the 3488A
mainframe. Six captive screws secure the cover.  To remove the cover:

• 

Remove all power cables, interconnections, and plug-in option cards from the 3488A.

• 

Turn the 3488A upside down. Loosen the six screws in the bottom cover. The screws are

captive in the bottom cover. Do not attempt to completely remove the screws.

• 

DO NOT ATTEMPT TO REMOVE THE BOTTOM COVER FIRST.  When the screws are

loosened, turn the entire unit upright. The top cover can now be lifted off.

• 

To reinstall the top cover, reverse the procedure.

STEP 2: Front Panel Removal

This procedure shows how to remove the front panel assembly from the mainframe.

• 

Depress the line power switch to the ON position.

• 

The front panel is now held captive in the slot in the bottom cover.  Gently pry the right side of

the bottom cover until the keyboard can be released from the slot. The keyboard assembly will
then swing out and clear the line power switch knob.

• 

Lift the front panel up and out. The cable that attaches the front panel to the Controller Board

has enough slack to allow the front to be removed. If it is desired to completely remove the front

Содержание 3488A

Страница 1: ...ABLE SERVICE RETURN USED RETURN CATEGORY ON SITE INVENTORY SCRAP PARTS SCRAP SERVICE CENTER SEE TEXT SEE TEXT AVAILABILITY PRODUCT S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL AUTHOR initials PRODUCT LINE AR ADDITIONAL INFORMATION AGILENT TECHNOLOGIES INC 2002 PRINTED IN U S A May 21 2002 Page 1 of 6 3488A 19 S E R V I C E N O T E Supersedes NONE 3488A Switch Control Mainframe Serial Numbers All US000...

Страница 2: ... for repair control module replacement Please call your local Contact Center for the Service Center nearest you 2 Customers can order an upgraded Control Module 03488 66541 and replace it themselves Please call your local Agilent Contact Center to place your order Control Module Replacement Instructions Step 1 Install updated 03488 66541 Control Board Follow 3488A Upgrade Instructions found below ...

Страница 3: ...dering irons solder suckers shelving and equipment stands Only one common ground should be provided at a workstation The workstation must be kept free of nonconductors No common plastics polybags cardboard cigarette or candy wrappers should be allowed There should not be rugs or carpet on the floor shelving or bench top Only proper containers should be used for shipping storing and transporting as...

Страница 4: ...ld only be removed if it is necessary to replace either the display assembly or the keyboard assembly To reinstall the front panel reverse the procedure Figure 1 Cable Removal STEP 3 Controller Board Access To access the controller circuit board remove the four screws in the shield over the board The shield will lift up and off The controller printed circuit board is now accessible for troubleshoo...

Страница 5: ...ided new Controller Board reverse the procedure Figure 3 Replace the Controller Board STEP 5 Verifying and Testing the Agilent 3488A Mainframe After replacing the Controller Board you need to verify and test the 3488A mainframe according to the following procedure Make certain that the provided new Controller Board the shield the front panel the top cover etc have been installed into the Agilent 3...

Страница 6: ... installed Control Board Follow this procedure to program the 3488A s Serial Number in the upgraded Control Board you have just installed 1 Send the following SCPI commands to the 3488A to download the serial number Enter the serial number of the 3488A you have upgraded DIAG INT ON DIAG SERIALNUMBER CN12345678 2 Then cycle power This will write CN12345678 to 3488A upgraded control board ...

Отзывы: