DC Output Adjustment
The function generator stores nine calibration constants related to
DC volts output. The constants are calculated from the adjustment value
entered. The calibration constants are stored following completion of
setup 59.
No calibration constants are stored if the procedures are aborted
at any other setup.
1
Use a DMM to measure the function generator dcV output voltage for
each setup in the following table. These adjustments use a HIGH Z
output termination.
Nominal Output
SETUP
DC Volts
Adjustment for:
50
- 8.0 Vdc
Negative offset gain
51
8.0 Vdc
Positive offset gain
52
0.0 Vdc
AM offset
53
0.0 Vdc
2 dB Pre-attenuator offset.
54
0.0 Vdc
4 dB Pre-attenuator offset.
55
0.0 Vdc
6 dB Pre-attenuator offset.
56
0.0 Vdc
8 dB Pre-attenuator offset.
57
0.0 Vdc
10 dB Pre-attenuator offset.
58
0.0 Vdc
12 dB Pre-attenuator offset.
59
0.0 Vdc
14 dB Pre-attenuator offset.
2
Use the CALIBRATE menu to adjust the displayed output voltage at
each setup to match the measured voltage and enter the value.
3
Perform the DC Function Offset Verification procedures on page 57.
Chapter 4 Calibration Procedures
DC Output Adjustment
76
Содержание 33120A
Страница 2: ......
Страница 10: ...8 ...
Страница 14: ...Contents 12 ...
Страница 15: ...1 Specifications 1 ...
Страница 20: ...18 ...
Страница 21: ...2 Quick Start 2 ...
Страница 37: ...3 Front Panel Menu Operation 3 ...
Страница 51: ...4 Calibration Procedures 4 ...
Страница 85: ...5 Theory of Operation 5 ...
Страница 103: ...6 Service 6 ...
Страница 114: ...112 ...
Страница 115: ...7 Replaceable Parts 7 ...
Страница 129: ...8 Schematics 8 ...
Страница 145: ...TOP SIDE 33120 66502 REV C 33120 66502 Component Locator Diagram 143 ...