121
TIMebase Commands
POSition
POSition
Command
:TIMebase:POSition <delay_time>
The :TIMebase:POSition sets the amount of time from the center of screen to
the trigger point of the waveform.
<delay_time>
If the horizontal time base is set between 50 s/div and 50 ms/div, the delayed
trigger time range is:
If the horizontal time base is set to less than 50 ms/div then the delayed trigger
time range is:
Query
:TIMebase:POSition?
The :TIMebase:POSition? query returns the value of the delayed trigger.
Returned Format
<delay_time><NL>
delay_time
6
div
±
time base setting
×
=
delay_time
14
div
–
time base setting to 1s
×
=
Содержание 3000 Series
Страница 1: ...Agilent Technologies Agilent 3000 Series Oscilloscopes Programmer s Reference ...
Страница 7: ...7 1 I O Module Installation and Configuration ...
Страница 13: ...13 2 Introduction to Programming ...
Страница 18: ...18 ...
Страница 19: ...19 3 Commands Quick Reference ...
Страница 27: ...27 4 Common Commands ...
Страница 33: ...33 5 Root Level Commands ...
Страница 40: ...40 ...
Страница 41: ...41 6 ACQuire Commands ...
Страница 47: ...47 7 BEEP Commands ...
Страница 50: ...50 ...
Страница 51: ...51 8 CHANnel n Commands ...
Страница 60: ...60 ...
Страница 61: ...61 9 COUNter Commands ...
Страница 65: ...65 10 DISPlay Commands ...
Страница 73: ...73 11 KEY Commands ...
Страница 78: ...78 ...
Страница 79: ...79 12 MASK Commands ...
Страница 88: ...88 ...
Страница 89: ...89 13 MEASure Commands ...
Страница 110: ...110 ...
Страница 111: ...111 14 SAVerecall Commands ...
Страница 117: ...117 15 TIMebase Commands ...
Страница 124: ...124 ...
Страница 125: ...125 16 TRIGger Commands ...
Страница 139: ...139 17 WAVeform Commands ...
Страница 162: ...162 ...
Страница 166: ...Index 166 ...