2200 TapeStation
57
Troubleshooting
6
Troubleshooting 2200 TapeStation ScreenPlex Images
Troubleshooting 2200 TapeStation ScreenPlex Images
Table 8
Overview troubleshooting 2200 TapeStation images
Image
Issue
Probable cause
Solution
Only the top and
bottom markers are
present in the
sample.
Either the sample is
not present or is too
faint to visualise.
Always vortex samples for 5 seconds before
placing in the 2200 TapeStation. ScreenTape can
detect <1 ng/µL
The bands appear
diagonally across the
lane.
A bubble was located
on top of the gel at
the gel/buffer
interface.
Always flick the ScreenTape before placing in the
2200 TapeStation.