background image

Содержание 10400B Series

Страница 1: ...6 For Safety information Warranties Regulatory information and publishing information see the pages at the back of this book Copyright Agilent Technologies Company 1998 1999 2006 All Rights Reserved Agilent 10400B Series Miniature Passive Oscilloscope Probes ...

Страница 2: ...See Probe parts supplied on page 4 for parts and accessories that were supplied with the probe If the contents are incomplete or damaged notify your Agilent Technologies Sales Office Inspect the instrument If there is mechanical damage or defect or if the probe does not operate properly or pass performance tests notify your Agilent Technologies Sales Office If the shipping container is damaged or ...

Страница 3: ... to solder and will not slip These probes include a variety of ground leads for attaching to many different ground points Probe packaging provides convenient storage and ordering information for replacement parts Other key features Narrow sharp replaceable probe tip Browser attachment with ultra sharp crown point Use this to probe the circuit in debug mode The crown point will not easily slip off ...

Страница 4: ...00B Series Probe Parts Item Description Quantity 1 Probe 1 2 Probe tip 2 3 Barrel insulator 2 4 Browser 1 5 General purpose retractable hook tip 1 6 Dual lead adapter 1 7 SMD clip 2 8 Spring ground 4 9 Alligator ground lead 1 10 Socketed ground lead 1 11 Screwdriver 1 12 Browser pogo pin spare 1 You can order additional quantities of these parts See the ordering information in this manual label cd...

Страница 5: ...3 2138 2 10440B probe tip qty 5 Yellow 5063 2171 2 10441B probe tip qty 5 Violet 5063 2172 2 10442B probe tip qty 5 Grey 5063 2139 3 Barrel insulator qty 4 4 Browser qty 1 5063 2115 5 General purpose retractable hook tip qty 2 5063 2135 6 Dual lead adapter qty 1 5063 2147 See below 7 SMD clip qty 5 5063 2149 8 Spring ground qty 4 9 Alligator ground lead qty 2 5063 2140 10 Socketed ground lead qty ...

Страница 6: ...ries Agilent 10467 68701 SMD microclips 2 yellow and 2 green for connection to SMT parts with lead spacings of 0 5 mm 0 020 inch to 0 8 mm 0 032 inch Agilent 5063 2143 probe tip to BNC adapter Agilent 0960 0301 BNC m to BNC f 50Ω feedthrough terminator Agilent N2768A qty 25 vertical mini probe socket ...

Страница 7: ...7 Agilent 10400B Series Miniature Passive Oscilloscope Probes Probe parts supplied Agilent N2766A qty 25 horizontal mini probe socket ...

Страница 8: ...robe length 2 meters 9 ns Safety Meets IEC1010 2 31 Pull strength BNC to probe tip 12 lb static pull Pollution degree 2 Normally only non conductive pollution occurs Occasionally however atemporary conductivity caused by condensation must be expected 10 MHz Frequency 200 300 MHz 500 MHz 3 MHz 100 MHz 300 5 100 30 MHz Excessive power dissipation area Acceptable probe operation area Max Voltage Inpu...

Страница 9: ...electronic etc with smaller transient overvoltages than installation category overvoltage category II Installation category overvoltage category II Local level appliances portable equipment etc with smaller transient overvoltages than installation category overvoltage category III WEEE Compliance Temperature operating 0 C to 55 C Humidity operating Up to 95 relative humidity at 40 C Altitude opera...

Страница 10: ...ible with oscilloscopes that have a 50 Ω input Probe compatibility Model Number Type of Probe Bandwidth Scope Probe System Division Ratio Input R Input C Scope Input R Comp Range Length in meters 10433B High impedance passive 300 MHz 10 1 10 MΩ 10 pF 1 MΩ 10 16 pF 2 10436B High impedance passive 300 MHz 10 1 10 MΩ 11 pF 1 MΩ 18 22 pF 2 10437B 50Ω Resistive Divider 1 GHz 1 1 50Ω n a 50Ω n a 2 10439...

Страница 11: ...ions 54830A Series 10433B 10439B 10440B MSO DSO601xA 10433B 10437B 10439B 10440B 10442B MSO DSO603xA 10433B 10437B 10439B 10440B 10442B MSO DSO605xA 10433B 10437B 10439B 10440B 10442B MSO DSO610xA 10433B 10437B 10439B 10440B 10442B Infiniium Oscilloscope Compatible 10400B Series Probe ...

Страница 12: ...erequirescleaning disconnectitfrom allpower sources and clean it with a soft cloth dampened with a mild soap and water solution Make sure the probe is completely dry before reconnecting it to a power source To connect oscilloscope probes 1 Attach the probe connector to the desired oscilloscope channel 2 Connect the probe to the circuit using any of the probing aids To operate the probe When operat...

Страница 13: ...obing the circuit You may unscrew the barrel insulator and remove it to accommodate tight probing situations The barrel insulator can be screwed to multiple positions With the barrel insulator screwed all the way on you may add a spring ground a supplied accessory to the probe tip to provide ground contact 2 With the barrel insulator screwed most of the way out the barrel insulator covers the grou...

Страница 14: ... Agilent Wedge Probe Adapter conductors are inserted into the space between the legs of the IC If you are using the Agilent Wedge Adapter refer to the Agilent Wedge Adapter User s Guide Connecting the Probe to an Agilent Wedge Adapter Agilent Wedge Adapters 16 signal Agilent Wedge Probe Adapters include 3 removable jumpers Agilent part number 1258 0141 for quantity of 1 jumper Agilent Model Number...

Страница 15: ...e measurement errors 1 Connect the probe from the appropriate oscilloscope channel to the front panel probe adjust signal terminal on the scope You can attach the probe with the hole in the probe body facing up to allow adjustment 2 Press Autoscale 3 Using the wide end of the supplied screwdriver tool set the low frequency compensation adjustment on the probe for the flattest pulse possible as sho...

Страница 16: ...robe tip 5 Connect the probe tip to the BNC to probe tip adapter 6 Connect the probe BNC to the appropriate oscilloscope channel 7 Press Autoscale then set the time base to 5 ns div 8 Set the sample mode to equivalent time and the of averages to 4 9 Rotate the two high frequency compensation resistors fully clockwise 10 Adjust the two high frequency compensation resistors alternately in small incr...

Страница 17: ...ogies shall pay shipping charges to return the product to the Buyer However the Buyer shall pay all shipping charges duties and taxes for products returned to Agilent Technologies from another country Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that in...

Страница 18: ...correctly performed or adhered to could result in personal injury Do not proceed beyond a Warning sign until the indicated conditions are fully understood and met C A U T I O N The Caution sign denotes a hazard It calls attention to an operating procedure practice or the like which if not correctly performed or adhered to could result in damage to or destruction of part or all of the product Do no...

Страница 19: ......

Страница 20: ...A Agilent Technologies Printed in the USA Manual Part Number 10433 92002 10433 92002 User s Guide ...

Отзывы: