HE/001132
Page 86 of 92
Document No. 001132, Revision 2.0
Agfa HealthCare
NodeID Livelink : 14140190
14 June, 2007
Attribute Name
Tag
VR
Value
Presence
of Value
Source
> Time
(0040, A122)
TM
EMPTY
Fixed
> Person Name
(0040, A123)
PN
EMPTY
Fixed
> Text Value
(0040, A160)
UT
EMPTY
Fixed
> Concept Code Sequence
(0040, A168)
SQ
EMPTY
Fixed
Acquisition Context Description
(0040, 0556)
ST
ANAP
(2)
Auto
Private attributes
Cassette orientation
(0019,10F5)
CS
LANDSCAPE or
PORTRAIT
ALWAYS
Auto
(1)
: Can only be available in case of XRDI. For details on the actual content, please refer to the XRDI Dicom
Conformance statement.
(2)
: Never present
6.1.1.4
MG
6.1.1.4.1
MG Image IOD
Table 6.1-6
: IOD of Created MG Image SOP Instances
IE
Module
Reference
Presence of Module
Patient
Patient Identification
Table 6.1-1
ALWAYS
Extended Patient
Table 6.1-1
If received from MWL
Study
General Study
Table 6.1-1
ALWAYS
Patient Study
Table 6.1-1
ALWAYS
General Series
Table 6.1-1
ALWAYS
DX Series
Table 6.1-5
ALWAYS
Series
Mammography Series
Table 6.1-7
ALWAYS
Equipment
General Equipment
Table 6.1-1
ALWAYS
Image
General Image
Table 6.1-1
ALWAYS
Image Pixel
Table 6.1-1
ALWAYS
Display Shutter
Table 6.1-5
When shutter is applied
DX Anatomy
Table 6.1-5
ALWAYS
DX Image
Table 6.1-5
ALWAYS
DX Detector
Table 6.1-5
ALWAYS
X-Ray acquisition dose
Table 6.1-5
When the XRDI Module is installed
X-Ray Collimator
Table 6.1-5
When the XRDI Module is installed
DX Positioning
Table 6.1-5
Mammography Image
Table 6.1-7
ALWAYS
VOI LUT
Table 6.1-5
ALWAYS
Acquisition Context
Table 6.1-5
ALWAYS
SOP Common
Table 6.1-1
ALWAYS