Service
Status
Further Meaning
Status Code
Behavior
*
Other Status codes
*
Other Dicom error codes result in the failure of the job.
Other warnings are not communicated to the user.
The behavior of the AE during communication failure is summarized in a table as follows:
Table 2.2-54: DICOM Command Communication Failure Behavior
Exception
Behavior
Timeout
If the sending fails, the system will retry automatically until the session is cleaned up.
Association aborted
If the sending fails, the system will retry automatically until the session is cleaned up.
The following table provides a description of the MPPS N-CREATE and N-SET request identifiers sent by the
MPPS AE. Empty cells in the N-CREATE and N-SET columns indicate that the attribute is not sent. An “X”
indicates that an appropriate value will be sent.
Table 2.2-55: MPPS N-CREATE/N-SET Request Identifier
Attribute
Tag
N-CREATE
N-SET
Specific Character Set
(0008, 0005)
X
X
Performed Procedure Step Relationship
Scheduled Step Attribute Sequence
(0040, 0270)
>Study Instance UID
(0020, 000D)
X
>Referenced Study Sequence
(0008, 1110)
From MWL or
user input
>>Referenced SOP Class UID
(0008, 1150)
From MWL or
user input
>>Referenced SOP Instance UID
(0008, 1155)
From MWL or
user input
>Accession Number
(0008, 0050)
From MWL or
user input
N/A
>Placer Order Number/Imaging Service Request
(0040, 2016)
From MWL or
user input
>Filler Order Number/Imaging Service Request
(0040, 2017)
From MWL or
user input
>Requested Procedure ID
(0040, 1001)
From MWL or
user input
>Requested Procedure Description
(0032, 1060)
From MWL or
user input
>Scheduled Procedure Step ID
(0040, 0009)
From MWL or
user input
>Scheduled Procedure Step Description
(0040, 0007)
From MWL or
user input
>Scheduled Protocol Code Sequence
(0040, 0008)
From MWL or
user input
>>Code Value
(0008, 0100)
From MWL or
user input
>>Coding Scheme designator
(0008, 0102)
From MWL or
user input
>>Coding Scheme Version
(0008, 0103)
From MWL or
user input