5.3
RADIOGRAPHIC OPERATION
RAD operation can be performed in the following modes:
One point control by selecting kV with AEC operations.
Two point control by selecting kVp and mAs independently.
Three point control by selecting kV, mA and Exposure Time
independently.
A typical RAD examination sequence is as indicated below:
1.
Make sure that the X-ray Tube is properly warmed-up.
2.
Position the patient for the examination.
3.
Select the “
DR Detector
” or “
Free Exposure
”
workstation
, and technique
parameters using the Generator controls on the software Console.
4.
Instruct patient to maintain the required position. Prepare the X-ray Tube
by pressing the Handswitch push-button to the “
Prep
” position and
maintain it until the “
Prep
” indicator is illuminated.
5.
Instruct patient to remain still and to hold his breath as required, then
make the X-ray exposure by pressing the Handswitch push-button fully
to the “
Exp
” position and maintain it throughout the exposure. The “
X-ray
On
” indicator will light and an alarm will sound during the exposure.
6.
When the exposure is finished, release the Handswitch push-button.
7.
Repeat the procedure if additional exposures are desired.
Содержание DX-D 300
Страница 1: ...DX D 300 U Arm User Manual Technical Publication 0171F EN 20181009 ...
Страница 2: ......
Страница 8: ...DX D 300 U Arm User Manual 0171F EN 20181009 iv ...
Страница 56: ...DX D 300 U Arm User Manual 48 0171F EN 20181009 This page intentionally left blank ...
Страница 100: ...DX D 300 U Arm User Manual 92 0171F EN 20181009 This page intentionally left blank ...
Страница 103: ......