User Manual
0171F EN 20181009
45
SECTION 3
START UP AND SHUTDOWN
3.1
START UP
The System should be powered by the same Room Electrical Cabinet where
the X-ray Generator is connected, that is, the whole System will be powered
from the same Electrical Cabinet.
To turn the System ON:
1.
Turn ON Room Electrical Cabinet Switch (the Emergency Stop not
activated).
2.
Press the X-Ray Control ON Button, the Green lamp turns ON.
3.
Press ON/OFF Switch located at the Control Box Door (the Emergency
Stop not activated) the Green lamp turns ON.
4.
Turn ON the NX Workstation.
Illustration 3-1
Start Up sequence
U-Arm Power
Room
Electrical
Cabinet
4
NX Workstation
3
Control Box
X-Ray Generator
1
X-Ray Generator Control
2
ON OFF