DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 7
06-2008 Type
5367/100
Agfa Company Confidential
2
Calibrations
Following calibrations have to be performed at the printer:
Adjustment
To be done ....
Details see....
Film calibration
•
During
maintenance.
•
After replacement of the Thermal
print head (TPH).
•
During the troubleshooting
procedure.
Internal densitometer
(CDM) calibration
•
When the density reproduction
check indicates to calibrate the
CDM.
•
During the troubleshooting
procedure.
Print Head (TH)
calibration
•
During the troubleshooting
procedure.
•
During the TH cleaning procedure.
Refer to section 3.2 in this chapter.
Smart Profile
Calibration
•
During the troubleshooting
procedure.
•
In certain cases during
maintenance. Refer to section 9,
Maintenance.
Compliance Film
Calibration
•
May be required as last step of an
Afssaps / EUREF calibration
procedure for a Drystar AXYS
Mammo printer.
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......