Quality Control Charts
D
RYSTAR
4500 U
SER
MANUAL
Drystar 4500: Determination of Operating Levels
Chart 1
Imager Type: __________
Serial #: _________________
Date ____________________
Film Type: ____________
Emulsion #: ______________
Densitometer
Internal: _________________ (default selection)
Day 1
Day 2
Day 3
Day 4
Day 5
Month
Day
Initials
Low Density
Average of 5 Values = operating (aim) level “Low Density”
Mid Density
Average of 5 Values = operating (aim) level "Mid Density"
High Density
Average of 5 Values = operating (aim) level “High Density”
Step 1:
Print QC Test images on five consecutive days. Record the optical densities
measurements in the tables below. After five days, average the values to determine
the operating (aim) levels for each of the parameters.
Step 2:
Copy the operating (aim) levels to Charts 2A/B ('Daily Density Control Chart')
2801 E EN UM Drystar 4500.fm Page 78 Monday, February 21, 2005 2:32 PM