Advanced operation (key-operator mode)
D
RYSTAR
4500 U
SER
MANUAL
Performing quality control (QC) tests
The following procedures must be performed daily, weekly or annually as
indicated.
The reason for performing quality control tests is to determine if any
significant image quality variation or deterioration has occurred which may
require corrective action. Comparing the results of the tests with the reference
values previously established does this.
This procedure allows the operator to take the necessary preventive actions
before any image quality loss can take place.
Performing the daily QC test
1
Turn on the Drystar 4500 and wait at least for 15 minutes. Refer to
on page
.
2
Press the Key-operator key to enter the Key-operator mode.
3
Press the Down key seven times, followed by the Confirm key to select ‘QC’.
The ‘Select input tray’ screen appears:
4
Press the Up/Down arrow keys to select the proper input tray, followed by the
Confirm key.
The Drystar 4500 will automatically print the QC Test image.
This test must be performed every day before any clinical film can be
processed.
2801 E EN UM Drystar 4500.fm Page 57 Monday, February 21, 2005 2:32 PM