89
2241A EN 20040210
Environmental requirements
Room temperature
20°C - 30°C
Maximum temperature change
0.5°C/min.
Relative humidity
10% - 80%
Magnetic field
Less than 5 Gauss; compliant with
EN 61000-4-8, Level 5
Sunlight exposure
Not to be operated in full sunlight
Physical emissions
Noise emission (sound power level according to DIN 45635 Part.27)
• During scanning
max. 65 dB(A)
• Standby
max. 46 dB(A)
Heat emission (at max. throughput with 35 cm x 43 cm image plates)
• Europe
0.8 kWh
• US
2730 BTU/hr
Cassette buffer capacity
10 cassettes of mixed sizes, both in input and output buffer
Performance
Throughput for high resolution
115 plates/h (size dependent)
Throughput for standard resolution
115 plates/h (size dependent)
Grayscale resolution
Data acquisition
12 bits/pixel
Output to processor
12 bits/pixel
Содержание CR 75.0
Страница 1: ...CR 75 0 Digitizer Reference manual CR 75 0 MUSICA...
Страница 36: ...36 2241A EN 20040210...
Страница 87: ...Equipment information sheet Appendix A...
Страница 91: ...ADC Compact cassette Appendix B...
Страница 101: ...101 2241A EN 20040210...
Страница 102: ...Printed in Belgium Published by Agfa Gevaert N V B 2640 Mortsel Belgium 2241A EN 20040210...