DD+DIS155.11E
Repair and Service
Troubleshooting
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Chapter 3.3 / 39 of 46
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential
CODE
Message
ARC4004
Shading calibration cannot be executed.
Meaning: Image processing software module reports an error during line
corrections.
The shading calibration data is not used successfully.
The image may be partly or completely uncorrected.
Reason 1: 'Optic Module' or 'PMT with Light Collector' replaced
without subsequent shading calibration.
Cure 1: Perform a shading calibration according to the service documentation.
Reason 2: Shading calibration file is corrupt.
Cure 2: Perform a shading calibration according to the service documentation.
Reason 3: The stored shading calibration data is not valid any more.
The Calibration line does not cover the complete width of the image to be
corrected.
Cure 3: Perform a shading calibration with the biggest used cassette format.
Refer to the service documentation.
Reason 4: Calibration line of this image plate type not available, invalid or could
not be loaded.
Cure 4: Perform a shading calibration with the image plate type, which has
generated the error.
Refer to the service documentation.
Reason 5: Software Bug. Preparation of the correction line is not successful.
The image is uncorrected.
Cure 5: Install newest software.
ARC4005
Software error in the image processing module
Meaning: Image processing software module indicates an error performing
calibration line corrections.
The image may be partly or completely uncorrected.
Reason 1: A bug in the software.
Cure 1: Restart digitizer to return IP into cassette and release cassette if
possible.
Remove jammed IP and cassette manually if restart not successful.
Install the newest digitizer software.
Содержание CR 10-X
Страница 8: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 84: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 104: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 116: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 148: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 202: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 228: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 236: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 283: ...1st Edition CR 10 X CR 12 X Type 5151 100 110 200 210 Order no DD DIS155 11E...