Table 7: Functional description of the positioning template.
Elements
Functional description
Red dot (1).
Used for positioning the positioning template. The red dot
must be situated under the anode side. When the anode is
on the left side, the red dot must be at the lower left hand
side. When the anode is on the right side, the red dot must
be at the upper right hand side.
Positioning
template
orientation
label (2)
This label shows the orientation towards the Heel effect.
POSITIONING
TEMPLATE
tube side
heel effect
Cassette / Phantom
Figure 7: Detail of the label.
Dosimeter
position
reference (3)
This is the point where the dosimeter is placed for the
determination of the exposure settings. So the dosimeter
position reference is used for measuring X-Ray doses.
Arrows and
lines (4)
These are used to align the positioning template together
with the collimator light lines in the longitudinal and
transverse side.
Holes to
position the
metal pins (5)
These are used to position different cassette sizes. Enter
metal pins and push the cassette so that the left and bottom
side touch the metal pins.
2 phantom
holders (6)
Used to position the phantom firmly on the positioning
template in combination with the phantom cuttings.
4 central
round cuttings
(7)
Is the holder for the light field indicator target when you
store the phantom with the light field indicator in the flight
case.
1.
Position the positioning template.
a) Switch on the collimator light.
b) Position the template on the table according to the following
specifications:
28
| Auto QC² | Auto QC² Hardware Components
2370C EN 20180330 1524