Full Leg Full Spine wall stand technical data
Table 9: Full Leg Full Spine wall stand technical data
Type
6005/100
Eelectrical connection operating
voltage
24V DC 2.0A
Dimensions
(without wall-mounted base)
Width: 735 mm (one handle)
Height: 1374 mm
Depth: 171 mm
Weight approx.
(without wall-mounted base)
25 kg
Front panel of the DR detector hold-
er attenuation
≤
1.2 mm Al
According to DIN EN 60601-1-3 with
100kV and HVL 3.6 mm Al, FDA 21
CFR § 1020.30 (n) with 100kV and
HVL 3.6 mm Al
Environmental requirements
Room temperature
10°C to 35°C
Maximum temperature change
0.5 °C/min.
Relative humidity
30% to 75%
Atmospheric pressure
70 kPa to 106 kPa
Relative altitude
3000 m to 0 m
Environmental requirements (storage)
Temperature
-20°C to +40°C
Environmental requirements (transport)
Temperature
-20°C to +40°C
Expected product lifetime
7 years
70
| Full Leg Full Spine DR Retrofit System | Technical Data
0326A EN 20211014 1141