Figure 15: Portrait orientation in the radiographic wall stand right loading
Orientation of DX-D 10C, DX-D 10G in the bucky
To avoid damage to the cable of the detector, there are restrictions on the
orientation of the detector when loading the bucky.
CAUTION:
Inserting DX-D 10C, DX-D 10G using other orientations than
described will damage the cable when closing the bucky or
when rotating the carrier.
Topics:
•
Orientation in the radiographic table
•
Orientation in the radiographic wall stand left loading
•
Orientation in the radiographic wall stand right loading
Orientation in the radiographic table
To use the detector in landscape orientation, insert the detector in landscape
orientation with the cable at lower right hand side.
To use the detector in portrait orientation:
1.
Insert the detector in landscape orientation with the cable at lower right
hand side.
2.
Rotate the detector in the bucky.
1
2
Figure 16: Portrait orientation in the radiographic table
Bucky | Bucky |