![AGFA 5430/110 Скачать руководство пользователя страница 86](http://html.mh-extra.com/html/agfa/5430-110/5430-110_user-manual_2866608086.webp)
DX-D 600 Technical Data
The technical data are provided in this chapter or in the User Manual of the
component.
Table 1: Environmental conditions for the X-ray system
Environmental Conditions (during storage and transport)
Temperature (ambient)
between -20 and 70 degrees Celsius
Humidity (non condensing)
between 10 and 70 % relative humidity
Atmospheric pressure
between 50 and 106 kPa
Environmental Conditions (during normal operation)
Temperature (ambient)
between 10 and 35 degrees Celsius
Humidity (non condensing)
between 30 and 75 % relative humidity
Atmospheric pressure
between 70 and 106 kPa
For overall system environmental conditions, the environmental conditions of
the DR Detector or image plate should be taken into account. Refer to the
related User Manual for environmental conditions for the DR Detector or
image plate. When using the DR Detector or image plate inside the bucky, take
into account that the temperature inside the bucky can be up to 5°C higher
than the temperature in the X-ray room.
86
| DX-D 600 | Technical Data
0240H EN 20181119 0940