CAUTION:
Damaged grid. Reduced image quality. Please handle the grids
with special care.
CAUTION:
When inserting the scattered radiation grids, it is essential that
the grid corresponds to the intended source-image-distance
(SID) to which the grid is focussed. Because of the focussing of
the grids, the tube unit must be centered onto the bucky.
CAUTION:
Power failure can cause an image being lost.
Connect the workstation and the digitizer to an uninterrupted
power supply (UPS) or an institutional standby generator.
• For product safety directions, refer to the User Manuals.