Installation
Installation and configuration is performed by an Agfa trained and authorized
service engineer. Contact your local support organization for more
information.
The unit shall not be installed in wet environments like emergency operation
rooms and operating theatre. The environment should be dust free and clean.
Environment of Use
The equipment is mainly for use in X-ray exposure rooms, hospital wards and
mobile medical examination vehicles. To use it in other places, consult your
sales representative or local Agfa dealer.
Precautions for installation:
• There must be no dripping water in the area.
• The environment must be free of harmful elements such as humid or
acidic air, air with a saline or sulfur content, poor ventilation, or unusual
air pressure or temperature.
• The equipment must not be placed at an angle or subjected to vibration or
shock (including when it is transported).
• The equipment must not be kept where chemical products are stored or
where gasses are generated.
• The site’s power supply must be of the correct voltage and frequency for
the equipment.
• The site must be connected to a fully grounded cable with sufficient
ground resistance to meet standard values.
Related Links
on page 14
DR 18M, DR 24M | Introduction to the DR Detector |
27
0380A EN 20180703 1429