DD+DIS303.03E
Installation
Planning
Edition 6, Revision 0
Classic E.O.S. / Classic E.O.S. CL
Chapter 14 / 33
(Type 5270 / 100 / 105)
9 Technical
Data
9.1 Electrical
data
Power connection
1N~ 230 – 240 V; 50/60 Hz (200 - 240 V)
Power consumption:
Standby
(room temperature ~ 20 °C)
0.45 kWh (1620 kJ)
during film processing
2.9 kW/h (10440 kJ/h)
Fuse protection
15 A / 16 A
Leakage current towards PE
< 3.5 mA
Main breaker
Upon machine installation it must be
ensured that either the mains
connector or an all-pole circuit
breaker for the installation on site is
located close to the machine and
easily accessible.
Safety regulations
Electrical installations in the installation
room must be incompliance with IEC 364
(VDE 0100 / 0107).
A GFI switch: (I
N
= 30 mA in
compliance with VDE 664) is
integrated in the machine.
9.2
Ambient and climatic conditions
Room temperature
min. 10 °C (50 °F), max. 30 °C (86 °F)
Room temperature min. 5 °C (41 °F) below the set
developer temperature
Relative humidity
min. 10 %, max. 80 %, no condensation
Ventilation
Avoid inhaling of chemical fumes. Make sure that
there is sufficient ventilation at the installation site
of the machine, i.e. an air exchange that is at
least ten times the room volume per hour.
maximum 2500 Lux
Light-tightness
The machine must not be opened
during the operation. It is light tight
only if all panels are correctly
mounted.
Acoustic test
ISO 7779
(airborne noise)
during standby
max. 35 dB (A)
during film cycle
max. 48 dB (A)