Installation Planning
DD+DIS303.03E
Chapter 14 / 22
Classic E.O.S. / Classic E.O.S. CL
Edition 6, Revision 0
(Type 5270 / 100 / 105)
8.4
Replenisher tanks for developer and fixer
Figure 21
Replenisher pump with filter
Replenisher tanks or Mixer
Replenisher hose
•
To be observed for the use of individual tanks or a mixer:
maximum
suction height
:
maximum
suction length
:
2 m (78.74 inch)
15 m (590.55 inch)
Replenisher tanks must not be installed inside the machine!
The customer cannot refill the tanks!
Level monitoring in the replenisher tanks
Four plugs are provided on the Control Board PCB1 for connection of the
replenisher supply for mixer, developer, fixer, and for the anti-algae solution.
A communication cable with 20 m (787.4 inch) an be ordered for the Mixer:
CM+9528030301