36
2231C EN 20040224
Re-erasing image plates with status ‘to be erased’
To re-erase one or more image plates which you have given the status ‘to be
erased’ via the ID station:
1
Check that the ADC Compact Plus is ready for operation:
•
the ADC Compact Plus must display the operator screen with ‘Ready’ or ‘Busy’
status.
•
the status indicator at the top of the ADC Compact Plus must be constant or
flashing green.
2
Place the cassettes in the input buffer.
The ADC Compact Plus will automatically erase the image plates. The display will
read:
When the ADC Compact Plus has erased the image plates, it displays the operator
main screen.
3
Remove the cassette(s) from the output buffer.