5. START-UP PROCEDURE 

5.1. Before start-up you should check that flowmeter

’s operation parameters 

specified  in  the  Passport  correspond  to  data  entered  in  the  flowmeter 
memory. 

5.2.  Parameters  of  PEA  are  entered  according  to  the  Operation  Manual, 

Sections 2.1 Part II. 

5.3. Processing of measurement results is adjusted according to the Opera-

tion Manual, Section 2.4, Part II. 

5.4.  External  devices  (analog  recorder,  modem,  etc.)  are  connected  to  the 

flowmeter,  communication  system  and  flowmeter  outputs  are  adjusted 
according to the instructions given in Part II of the Operation Manual. 

5.5.  Current  date  and  time  are  checked  and  set  if  necessary  (see  Section 

2.5, Part II of the Operation Manual).  

5.6.  The  device  should  be  switched  into  the  OPERATION  mode  (a  jumper 

should be removed from terminal J4). Terminal J4 should be sealed. If 
necessary, two fixing screws are sealed from the front panel side. 

If the  parameters are set  correctly, the flowmeter displays flowrate 

measurement results. 

5.7. Requirements  for  mounting  location  and  operating  conditions  specified 

in this operational documentation are based on the most typical factors 
affecting flowmeter performance. 

The  external  factors  that  cannot  be  foreseen,  evaluated  or  tested 

by the manufacturer during designing may exist or appear on site. 

In this case such factors should be eliminated or flowmeter should 

be placed in other location. 

Содержание UF Series

Страница 1: ...ULTRASONIC FLOW METER AFLOWT UF with cut in converters VERSION UF 5xx d INSTALLATION MANUAL ISO 9001 2015 ...

Страница 2: ... for four beam scanning 6 3 2 Length requirements for straight pipe runs 7 3 3 Requirements for placing the Secondary measuring converter 7 4 MOUNTING THE FLOWMETER 8 4 1 Mounting the Primary Converter 8 4 2 Mounting the Secondary Measuring Converter 8 4 3 Wiring the Flowmeter 9 5 START UP PROCEDURE 11 6 DISMOUNTING 12 APPENDIX A Relative length of straight pipe runs 13 APPENDIX B Report on mounti...

Страница 3: ...WT UF with cut in converters Version UF 5xx d Operation manual part 1 and 2 LIST OF ABBREVIATIONS CS DN MC CP PEA Secondary measuring converter Nominal diameter Measuring cut Primary converter Electro acoustic converter NOTE Words in the text marked in bold for example Settings cor respond to the items displayed on the flowmeter s screen CAUTION It is highly recommended to study the Operation Manu...

Страница 4: ...lectronic components when the flowmeter is powered up To use defective electronic devices and electric tools or use them with out proper grounding 1 5 Prior to connection of the flowmeter to the electric mains the CS module is to be connected to the protective earthing trunk line neutral earth ing CAUTION Prior to connecting the protective earthing trunk line make sure that the voltage is off 2 MO...

Страница 5: ...ocation if applicable is at the lower or uprising pipeline section Fig 1 Recommended locations of the CP 3 1 1 2 The difference in the internal diameters of the pipeline and a measuring cut MS of the CP at a link up point should not exceed 0 05 DN for straight measuring cuts and 0 1 DN for CPs of U elbow type 3 1 1 3 It is not recommended to install CP so that the PEA was in the upper or lower cir...

Страница 6: ...m PEA2 1 PEA2 2 pair 2nd beam Fig 3 Placing the PEA pairs on chords single beam sounding 3 1 4 Placing the PEA for four beam scanning In case of four beam scanning the PEA pairs are mounted on the CP along four chords placed horizontally Fig 4 The order of beams can be reverse PEA1 1 PEA1 2 pair 1st beam PEA2 1 PEA2 2 pair 2nd beam PEA3 1 PEA3 3 pair 3rd beam PEA4 1 PEA4 2 pair 4th beam Fig 4 Plac...

Страница 7: ... in upstream position and the length of the straight pipe runs shall be de fined in view of this condition 3 2 2 In case of mounting the straightening vane into the pipeline before the measuring section Appendix D reduction of straight pipe run length is possible at the inlet of the CP up to two times 3 2 3 For the flowmeter with PEAs mounted along the axis of the straight CP of U elbow type the r...

Страница 8: ...or 2 companion flange with the pipe adaptor 3 CP Fig 5 Assembly welded structure for mounting the flanged CP into the pipe line 4 1 4 A pipe run of the required length is cut at the selected liquid free pipe line section and a structure with a flanged CP is welded instead The arrow on the CP shall point to the flow direction or forward flow direction in case of bidirectional measurements CAUTION I...

Страница 9: ... network cables are fixed on the wall wherever possible The network cable is laid separately at a distance of 30 cm as a minimum from other cables To avoid mechanical damages it is rec ommended to place all cables in metal tubes or sleeves IT IS FORBIDDEN to lay CS PEA cables and external signal ca bles within proximity to power circuits With presence of high level elec tromagnetic interference fo...

Страница 10: ...electric installations Do not use the light ning protection system for grounding since in could result in device s failure As a grounding wire you must use a copper ground wire between the flowmeter and grounding device If mechanically protected its cross section must be minimum 2 5 mm2 Minimum 4 mm2 cross section is required for installations without protection The grounding conductor is connecte...

Страница 11: ...t date and time are checked and set if necessary see Section 2 5 Part II of the Operation Manual 5 6 The device should be switched into the OPERATION mode a jumper should be removed from terminal J4 Terminal J4 should be sealed If necessary two fixing screws are sealed from the front panel side If the parameters are set correctly the flowmeter displays flowrate measurement results 5 7 Requirements...

Страница 12: ...eline section where they are located ensure there is no pressure in the pipeline and drain the liquid Disconnect cables leading to the CS PEAs Dismount the CP and CS 6 2 Before dispatching the flowmeter with the CP for calibration or repair clean the measuring cuts and emitting surfaces of PEAs from sediment and liquid residues after dismounting ...

Страница 13: ...type installation different scheme of sounding depend ing on local hydraulic resistance types Table A 1 Type of local hydraulic resistance Relative length of a straight run N minimum single beam diame tral sounding two beam chord sounding four beam chord sounding 1 2 3 4 10 10 10 3 1 1 10 10 10 3 1 1 10 10 10 3 1 1 10 10 10 10 10 10 ...

Страница 14: ... 1 In case of various types of hydraulic resistance in the pipeline the length of a pipeline straight run related to the resistance nearest to the PEAs should be no less than the value specified in the table and the distance from the PEAs to the other hydraulic resistances should be no less than the value specified in the table for hydraulic resistance of this type ...

Страница 15: ..._________________________ 2 Site characteristics DN of measuring cut _______ mm PEA installation diagram ____________ negative flow ________ Type of sounding _________ beam Table B 1 Type of hydraulic resistance Length of a straight run m To PEA1 1 2 1 3 1 4 1 After PEA1 2 2 2 3 2 4 2 3 Parameters of the medium 3 1 Kind of liquid _____________________________________________ 3 2 Temperature of liq...

Страница 16: ...________________________________________ Representative of start up organization ____________________ ______________________ Signature name and initials ___ ______________ 20__ Customer representative ____________________ ______________________ Signature name and initials ___ ______________ 20__ Calibrator ____________________ ______________________ Signature name and initials LS ___ _____________...

Страница 17: ...00 84 00 85 00 86 00 87 00 88 00 89 00 90 00 91 00 92 00 93 00 94 00 95 00 96 00 97 00 98 00 99 00 100 0 101 0 102 0 103 0 104 0 0 4137 0 4083 0 4030 0 3979 0 3929 0 3880 0 3832 0 3785 0 3740 0 3695 0 3651 0 3608 0 3566 0 3525 0 3485 0 3446 0 3407 0 3370 0 3333 0 3297 0 3261 0 3227 0 3193 0 3159 0 3127 0 3095 0 3064 0 3033 0 3003 0 2973 0 2944 0 2916 0 2888 0 2861 0 2834 105 0 106 0 107 0 108 0 10...

Страница 18: ...ore rare ly at the periphery d holes have chamfers from the flow inlet site D internal diameter of the pipeline where the straightening vane is inserted Fig D 1 Design of type A flow straightening vane 2 To lower weight and material quantity a type B straightening vane Fig D 2 can be used it is manufactured according to the following rules a tubes are inserted in the plate holes b tube length is e...

Страница 19: ... D 0 078 D 6 0 134 D 0 225 D 7 0 156 D 0 381 D 8 0 252 D 0 9 0 255 D 0 146 D 10 0 288 D 0 288 D 11 0 396 D 0 12 0 400 D 0 151 D 4 The straightening vane is installed in the pipeline at a distance of 1 2 pipeline s DN from the last source of hydraulic resistance with respect to the flow direction When installing the straightening vane the length of a straight run before the PEA is determined as the...

Отзывы: