Prosthesis Settings Tab
12
Ignore EMG spikes
If the patient is facing a difficulty with EMG spikes - for example while exercising -
you can activate the “ignore EMG spikes” function. By doing this the prosthesis will
ignore short spikes of EMG of the length specified by the adjustable slider.
WARNING
: Keep in mind that this function will create a delay in the prosthesis
response time, as the initial part of the signal is ignored.
Auto-grasp
Auto-grasp is a function for preventing slippage of the object being held. Once
the fingers close around an object, it is detected and a constant small force is
applied to prevent dropping it. If the object starts slipping from the hand, the
fingers will close to prevent it falling. The force of the fingers can be adjusted by
changing the Force slider.
WARNING
: Keep in mind that using this function increases the battery drain
drastically.