Use the Built-in Instrument
Check the connection of the power cord
and turn the instrument switch to "Internal
Instrument".
Press the power button, the button light will
turn green; (Note: The system restart interval
is 30s, please ensure a fully powered off
before the system reboot!).
The yellow light of the signal indicator turns
on and
fl
ashes, indicates that the internal test
module is booting up.
①
②
③
④
After 20 seconds, the green light is turned on,
indicates that the system has be started up
successfully and can be used normally.
System Power On
Website
Please visit
www.generaltest.com
for related documentation and technical support.
Send email to
for problem report, the technical engineer will provide a solution
accordingly and timely.
Contact us directly by (+86) 755 2940 3544.
Hotline
06
07
Trouble Description
Possible Reasons
Solutions
Client software can't find the
RN200A
The RN200A does not startup
normally or the LAN cable is not
connected correctly
The computer and the RN200A are
not configured in the same local
area network
Observe the signal indicator
is abnormal or not and check
the LAN cable connection
Configure the system
computer and RN200A to
the same local area network
Turn table reports error after
RN200A is restarted
Insufficient power shut off for the
turntable controller
Fully power off and wait 30s
before turning power on
Table 5 Trouble Shooting
Trouble Shooting