OPERATION 

Please make sure that you have already read and fully understand the WARNING section on page 1. 

Making Measurements with the AC Current Probe Model SR751/SR752 

• 

Connect the black lead of the current probe to “Common” and the red lead to the AC voltage input on  
your  DMM  or  other  voltage-measuring  instrument.  The  AC  current  probe  has  a  ratio  of  1000:1.  This 
means that for 1000A AC in a conductor around which the probe is clamped, 1V AC will come out of 
the  probe  leads  to  your  DMM  or  instrument.  The  output  is  1mV  AC/A  AC.  Select  the  range  which 
corresponds to the measured current. If the current magnitude is unknown, start with the highest range 
and  work  down  until  the  appropriate  range  and  resolution  are  reached.  Clamp  the  probe  around  the 
conductor. Take the reading on the meter and multiply it by 1000 to obtain the measured current (e.g, 
260mV reading = 260 x 1000 = 260,000mA or 260A) 

• 

For best accuracy, avoid if possible, the proximity of other conductors which may create noise. 

Tips for Making Precise Measurements 

• 

When  using  a  current  probe  with  a  meter,  it  is  important  to  select  the  range  that  provides  the  best 
resolution. Failure to do this may result in measurement errors. 

• 

Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air 
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for 
power measurement. 

MAINTENANCE 
Warning:

 

• 

For maintenance use only original replacement parts. 

• 

To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so. 

• 

To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into 
the probe 

Cleaning: 

To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times. 
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine 600) 
to avoid scratching the jaw, then gently clean with a soft oiled cloth. 

REPAIR AND CALIBRATION 

You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure 
that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the 
outside of the shipping container.  

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

 Instruments 

 

15 Faraday Drive 

 Dover, NH 03820 USA 

(800) 945-2362  (Ext. 360) • (603) 749-6434  (Ext. 360) • [email protected] 

(Or contact your authorized distributor) 

NOTE: All customers must obtain a CSA# before returning any instrument. 

TECHNICAL AND SALES ASSISTANCE 

If you are experiencing any technical problems, or require any assistance with the proper use or application 
of this instrument, please contact our technical hotline: 

(800) 343-1391 • (508) 698-2115 • [email protected] 

LIMITED WARRANTY 

The current probe is warranted to the owner for a period of two  years from the date of original purchase 
against defects in manufacture. This limited warranty is given by AEMC

®

 Instruments, not by the distributor 

from  whom  it  was  purchased.  This  warranty  is  void  if  the  unit  has  been  tampered  with,  abused  or  if  the 
defect is related to service not performed by AEMC

®

 Instruments.  

Full warranty coverage and product registration is available on our website at: 

www.aemc.com/warranty.html

 

Please print the online Warranty Coverage Information for your records.

 

Отзывы: