ELECTRICAL SPECIFICATIONS
Current Range:
0.1 to 1000A AC, continuous cycle @
1kHz
Output Signal:
1mV AC/A AC (1V AC at 1000A)
Accuracy and Phase Shift*:
Accuracy:
0.1 to 1A:
2% Reading ± 3µV
1 to 10A:
1% Reading
10 to 100A:
0.5% Reading
100 to 1200A:
0.3% Reading
Phase Shift:
0.1 to 1A:
Not Specified
1 to 10A:
2°
10 to 100A:
1°
100 to 1200A:
0.7°
*Reference conditions: 23°C±3°K, 20 to 75% RH, 48
to 65Hz, external magnetic field <40A/m, no DC
component, no external current carrying conductor,
test sample centered. Load impedance 1M
.
Overload:
1200A for 40mn on, 20mn off
Accuracy
: Per IEC 185