26
AC/DC Current Probes MR417 & MR527
Pulse Response (MR417 and MR527)
Rise time from (10 to 90) %: ≤ 11 μs
Fall time (from 90 to 10) %: ≤ 11 μs
AC noise on output: ≤ 3 mV or 0.3 A
Delay time at 10 %: ≤ 10 μs
Square wave response curves
4.3 OPERATING LIMITS (MR417 AND MR527)
•
In DC: 3000 A permanent
•
In AC: 1000 A permanent up to 1 kHz
•
from 1 kHz, IMAX = 1000/f (kHz)
•
Conductor temperature: ≤ 194 °F (90 °C), 230 °F (110 °C) peak
•
Temperature of the jaws: ≤ 176 °F (80 °C)
Curve of derating vs frequency
1
10
0
100
200
300
600
400
500
700
800
1000
900
Frequency
(kHz)
Current (A)