AC/DC Current Probe Models MR411 & MR521
- 7 -
Operating Relative Humidity:
10-35°C: 90% ±5% RH (without condensation)
40-55°C: 70% ±5% RH (without condensation)
RH Influence:
10-35°C: 90% RH at reference
Temperature:
0.5% (MR411)
0.1% (MR521)
Altitude:
Operating: 0 to 2000m
Non-operating: 0 to 12,000m
Zero Adjustment:
Automatic zero (± 10) A by simple push button
increments of 25 to 40mA. Red LED goes on at push of the button and
turns off when zero is reached; no need to the hold button down.
Case protection:
IP30 per IEC529
Drop Test:
1.0m on 38mm of oak on concrete; test according to IEC 1010
Mechanical Shock:
100G, test per IEC 68-2-27
Vibration:
Test Per IEC 68-2-6,
Frequency Range:
5 to 15Hz, amplitude: 1.5mm
15 to 25Hz, amplitude 1mm
25 to 55Hz, amplitude: 0.25mm
Handle:
UL94 V0
Jaws:
UL94 V0
Color:
Dark gray with red jaws
Output:
Double/reinforced insulated 5ft (1.5m) with safety 4mm banana plug